Note that despite being properly calibrated
it has been my frequent experience that different pH meters will prod
uce
significantly different pH readings on the same solutions. This is
especially true of lower quality electrodes because they can respond to
other 'species' which are electrochemically similar to hydrogen ions, such
as sodium. For this reason a hydroponic grower can get pH readings differing by more
than 0.5 units when testing a nutrient sample with two different pH meters
that have been correctly calibrated. Note that pH indicators cannot give
false readings with colorless samples because they will only respond to
hydrogen ions.
Storage
of
pH Electrodes
For a pH meter with a single combination
electrode to work properly there must be an electrical connection between
the electrode filling solution (usually potassium chloride – which is
electrically conductive) and the sample solution. This is achieved via the
use of a porous frit or wick (Fig 1.8a) in the glass wall separating the
inner filling solution chamber from the sample. Both these devices allow
the free (very slow) flow of the filling solution into the sample
solution.
The
most common cause of pH meter failure is due to blockage of this porous
frit or wick. This is usually caused by dry storage of the electrode or
leaving the electrode immersed for too long in concentrated solutions
(e.g. concentrated nutrient). Dry storage causes dehydration of both the
glass electrode and precipitation of salts within the interstices of the
frit itself (Fig 1.8b). The consequences are reading drift, slow response
times and decreased sensitivity (loss of accuracy and precision). To avoid
these problems the electrode tip must be permanently stored (Fig 1.7) in a
specially formulated storage solution such as Flairform's
pH Electrode Storage
solution.
NOTE: pH buffers 7.0 or 4.0 and
distilled water are not suitable for storing pH electrodes as they can
upset electrode calibration.
Cleaning
pH Electrodes
The ‘frit’ and glass tip of pH meter
electrodes invariably become contaminated with impurities (Fig 1.8b)
causing symptoms such as reading drift, slow response times and decreased
sensitivity (loss of accuracy and precision). Contamination may be so
severe that re-calibration is not possible without prior cleaning.
Minimise this problem by filtering
potentially problem (i.e. greasy or highly turbid) water samples prior to
performing a pH measurement. Avoid placing electrodes in oily solutions as
they can “coat” the electrode. Further, avoid measuring (or leaving
electrodes for too long) in harsh chemicals. For example, highly caustic
(e.g. pH Up, silica additives) or concentrated solutions (e.g. raw
nutrient). These can damage electrodes and/or increase the need for
electrode maintenance.
Regular cleaning will maximize electrode
performance and ensure a longer working life. Some suitable cleaners are:
- Organic contamination: Use mild
liquid detergent, acetone*, methylated spirit or toluol*. If toluol is
necessary, rinse subsequently with acetone or methylated spirits and
finally water.
- Inorganic contamination: Use
dilute nitric or hydrochloric acid.
In hydroponics, because both organic and
inorganic contaminations tend to co-occur, it is best to use an
all-purpose cleaner such as Flairform’s
Electrode Cleaner.
* Note that some cleaners may attack
electrode hardware - particularly plastic. Therefore ensure to check with
the electrode’s manufacturer before proceeding.
Next section...